Agilent Technologies 8163A Programming Manual page 78

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings
command:
syntax:
description:
parameters:
Measurement Running
Measurement Stopped
NO TE
NO TE
NO TE
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
78
N OT E
Setting parameters for the logging function sets some parameters,
including hidden parameters, for the stability and MinMax functions and
vice versa. You must use the
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing command to
set parameters before you start a logging function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe command.
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing<wsp><data points>,
<averaging time>[NS|US|MS|S]
Sets the number of data points and the averaging time for the logging data acquisition
function.
Data Points:
Data Points is the number of samples that are recorded before the
logging mode is completed. Data Points is an integer value.
Averaging time:
Averaging time is a time value in seconds.
There is no time delay between averaging time periods. Use
":SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility?" on
page 81 if you want to use delayed measurement.
1
2
3
If you specify no units for the averaging time value in your command, seconds are used as
the default.
See ":SENSe[n][:CHANnel[m]]:FUNCtion:STATe" on page 82 for information on start-
ing/stopping a data acquisition function.
See ":SENSe[n][:CHANnel[m]]:FUNCtion:RESult?" on page 81 for information on accessing
the results of a data acquisition function.
See "Triggering and Power Measurements" on page 117 for information on how triggering
affects data acquisition functions.
none
sens1:func:par:logg 64,1ms
All power meters and return loss modules
Can only be sent to master channel, slave channel is also affected.
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing?
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing?
Returns the number of data points and the averaging time for the logging data acquisition
function.
none
Returns the number of data points as an integer value and the averaging time, t
float value in seconds.
sens1:func:par:logg?
All power meters and return loss modules
Can only be sent to master channel, slave channel parameters are identical.
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Measurement Functions – The SENSe Subsystem
Averaging Time
4
5
6
7
+64,+1.00000000E-001<END>
8
9
t
avg
, as a