Agilent Technologies 8163A Programming Manual page 89

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Functions – The SENSe Subsystem
command:
syntax:
description:
parameters:
response:
example
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example
affects:
command:
syntax:
description:
parameters:
response:
example
affects:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength?
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength?[<wsp>MIN|MAX|DEF]
Inquires the current sensor wavelength.
none
Also allowed are:
The wavelength as a float value in meters.
sens1:pow:wav? → +1.55000000E-006<END>
All power meters and return loss modules
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
Overwrites the current calibration values with the factory-set calibration settings. See
":SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance" on page 89 and
":SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination" on page 90 for
information on calibrating your return loss module.
none
none
sens1:ret:cal:fact
All return loss modules
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
Start the calibration and save the calibration values for a defined reflectance reference
measurement. See ":SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]" on
page 91 for information on setting the return loss value of your reference reflector.
none
none
sens1:ret:cal:coll:refl
All return loss modules
Measurement Operations & Settings
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is
half the sum of, the minimum programmable value and
the maximum programmable value
89