Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
106
[:SOURce[n]][:CHANnel[m]]:WAVelength[:CW|:FIXED]
[:SOURce[n]][:CHANnel[m]]:WAVelength[:CW|:FIXED]<wsp><value>
[PM|NM|UM|MM|M]
Sets the absolute wavelength of the output.
Any wavelength in the specified range (see the specifications in the appropriate User's
Guide).
The programmable range is larger than the range specified in the User's Guide. The pro-
grammable range is set individually for each instrument when it is calibrated during pro-
duction.
Also allowed are:
The default units are M.
none
sour2:wav 1550NM
All tunable laser and DFB source modules
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Signal Generation – The SOURce Subsystem
MIN: minimum wavelength value
MAX: maximum wavelength value
DEF: This is not the preset (*RST) default value but is half the
sum of, the minimum wavelength value and the maximum wave-
length value