Agilent Technologies 8163A Programming Manual page 62

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Instrument Setup and Status
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
62
:STATus:QUEStionable:ENABle:LEVel1
:STATus:QUEStionable:ENABle:LEVel1<wsp><value>
Sets the bits in the Questionable Status Enable Summary Mask (QSESM) that enable the
contents of the QSESR for slots 15 - 17 of the Agilent 8166A Lightwave Measurement Sys-
tem to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR for slots 15 - 17
of the Agilent 8166A Lightwave Measurement System to affect bit 7 of the Status Byte.
The bit value for the QSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
none
stat:oper:enab:level1 128
:STATus:QUEStionable:ENABle:LEVel1?
:STATus:QUEStionable:ENABle:LEVel1?
Returns the QSESM for the QSESR for slots 15 - 17 of the Agilent 8166A Lightwave Mea-
surement System
none
The bit value for the QSESM as a 16-bit signed integer value (0 .. +32767)
stat:oper:enab:level1? → +128<END>
:STATusn:QUEStionable[:EVENt]?
:STATusn:QUEStionable[:EVENt]?
Returns the questionable status of slot n - the Questionable Slot Status Event Register
(QSSER).
none
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
Mnemonic
8-15
Not used
7
Slot n: Duty cycle has been out of range
6
Slot n: ARA has been recommended
5
Slot n: Module has been out of specification
4
Slot n: Module has settled unsuccessfully
3
Slot n: Laser protection has been on
2
Slot n: Temperature has been out of range
1
Slot n: A Zeroing operation has failed
0
Slot n: Excessive Value has occurred
Every nth bit is the summary of slot n.
stat1:oper? → +0<END>
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Status Reporting – The STATus Subsystem
Decimal Value
0
128
64
32
16
8
4
2
1

Advertisement

Table of Contents
loading