Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
108
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency<wsp><value>
[THZ|GHZ|MHZ|KHZ|HZ]
Sets the frequency difference used to calculate a relative wavelength. The output wave-
length is made up of the reference wavelength and this frequency difference.
The default units for frequency are Hertz.
The output wavelength (λ) is set from the base wavelength (λ
(df). The formula for calculating the output wavelength is:
where c is the speed of light in a vacuum (2.990 x 10
The frequency difference is a float value in Hz.
none
sour2:wav:freq -10THZ
All tunable laser and DFB modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency?
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency?
Returns the frequency difference used to calculate a relative wavelength.
none
Returns the frequency difference as a float value in Hz.
wav:freq?
-1.00000000E+013<END>
All tunable laser and DFB modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:REFerence?
[:SOURce[n]][:CHANnel[m]]:WAVelength:REFerence?
Returns the reference wavelength (λ
none
The wavelength as a float value in meters.
sour2:wav:ref?
All tunable laser and DFB modules
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Signal Generation – The SOURce Subsystem
c ( )
λ
----------------------------- λ
=
(
(
λ
)
df
+
o
).
0
+1.5500000E-006<END>
) and the frequency offset
0
)
o
c
8
-1
ms
)