Agilent Technologies 8163A Programming Manual page 118

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings
Hardware
Trigger Rearming
Triggering
trig:outp
trig:outp:rearm
DISabled
-
AVGover
ON
MEASure
ON
AVGover
OFF
MEASure
OFF
command:
syntax:
description:
parameters:
NO TE
NO TE
response:
example:
118
Table 8
Generating Output Triggers from Power Measurements
Software Triggering
init:imm
An output trigger will never be generated.
An output trigger is generated for every new power measurement when the averag-
ing time period finishes.
An output trigger is generated for every new power measurement when the averag-
ing time period begins.
An output trigger is generated when the averaging
time period of the first power measurement finishes.
A further hardware output trigger cannot be gener-
ated until you send trig:outp:rearm.
An output trigger is generated when the averaging
time period of the first power measurement begins.
A further hardware output trigger cannot be gener-
ated until you send trig:outp:rearm.
:TRIGger
:TRIGger<wsp>NODEA|1|NODEB|2
Generates a hardware trigger.
1 or NODEA:
Is identical to a trigger at the Input Trigger Connector.
2 or NODEB:
Generates trigger at the Output Trigger Connector.
A hardware trigger cannot be effective in the DISabled triggering mode but can be
effective in DEFault, PASSthrough or LOOPback triggering modes, see
":TRIGger:CONFiguration" on page 122 for information on triggering modes.
":TRIGger" on page 124 describes the :TRIGger command for advanced users
using ":TRIGger:CONFiguration:EXTended" on page 124.
none
trig 1
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
init:cont
MINMax
Triggering - The TRIGger Subsystem
Data Acquisition Functions
sens:func:stat
LOGGing|STABility
Applies for all subsequent
data acquisition functions.
Applies for all subsequent
data
acquisition functions.
An output trigger is generated
for every new power measure-
ment when the averaging time
period finishes. Applies for all
subsequent data acquisition
functions.
An output trigger is generated
for every new power measure-
ment when the averaging time
period begins. Applies for all
subsequent data acquisition
functions.