Measurement Functions – The SENSe Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold<wsp><mode>,
<threshold value>[PW|NW|UW|MW|Watt|DBM]
Sets the start mode and the threshold value.
ABOVe:
BELow:
IMMediately:
Threshold Value:
none
sens1:func:thr IMM,20nw<END>
All HP 8153A Lightwave Multimeter series power meters and the HP 81534A Return Loss
module
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold?
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold?
Returns the start mode and the threshold value.
none
ABOV:
Function starts when power is above the threshold value.
BEL:
Function starts when power is below the threshold value.
IMM:
Function starts immediately.
Threshold Value:
A float value in Watts or dBm.
→
sens1:func:thr?
All HP 8153A Lightwave Multimeter series power meters and the HP 81534A Return Loss
module
:SENSe[n]:[CHANnel[m]]:POWer:ATIMe
:SENSe[n]:[CHANnel[m]]:POWer:ATIMe<wsp><averaging time>[NS|US|MS|S]
Sets the averaging time for the module.
The averaging time as a float value in seconds.
If you specify no units in your command, seconds are used as the default.
none
sens1:pow:atim 1s
All power meters and return loss modules
Can only be sent to master channel, slave channel is also affected.
Measurement Operations & Settings
Function starts when power is above the threshold value.
Function starts when power is below the threshold value.
Function starts immediately.
A float value in Watts or dBm.
IMM,+2.00000000E-008<END>
83