Error Codes
New/Old/Standard
Standard
Standard
New
-200 to -299 Execution Errors
Standard
New
Old
Old
Old
Old
Old
Standard
Standard
Standard
194
Table 17
Overview for Supported Strings
Error
Number
-181
-183
-185
-200
-205
-211
-212
-213
-220
-221
NO TE
If error -221 is returned after you try to start a wavelength sweep, one
of the following cases of sweep parameter inconsistency has
occurred:
• Continuous Sweep mode AND λ Start is less than λ Stop.
• Continuous Sweep mode AND Sweep Time is too short. Adjust
Sweep Speed, λ Start, or λ Stop.
• Continuous Sweep mode AND Sweep Time is too long. Adjust
Sweep Speed, λ Start, or λ Stop.
• Continuous Sweep mode AND Trigger Frequency is too high.
Adjust Step Size. Trigger Frequency is the Sweep Speed divided by
the Step Size.
• Stepped Sweep mode AND Lambda Logging Enabled.
• Continuous Sweep mode AND Lambda Logging Enabled AND
Output trigger mode not set to STFinished (Step finished).
• Continuous Sweep mode AND Lambda Logging is Enabled AND
Modulation Source is not set to OFF.
• Continuous Sweep mode AND Lambda Logging is Enabled AND
Sweep Cycles is not set to 1.
• Continuous Sweep mode AND Coherence Control is Enabled.
-222
-223
-224
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
String
"Invalid outside macro definition"
"Invalid inside macro definition"
"Subop out of range"
"Execution error"
"Powermeter not running"
"Trigger ignored"
"Arm ignored"
"Init ignored"
"Parameter error"
"Settings conflict"
"Data out of range"
"Too much data"
"Illegal parameter value"
GPIB Error Strings