Agilent Technologies 8163A Programming Manual page 50

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Instrument Setup and Status
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
50
*OPC
*OPC
The instrument parses and executes all program message units in the input queue and
sets the operation complete bit in the standard event status register (SESR). This com-
mand can be used to avoid filling the input queue before the previous commands have fin-
ished executing.
The following actions cancel the *OPC command (and put the instrument into Operation
Complete, Command Idle State):
• Power-on
• the Device Clear Active State is asserted on the interface.
• *CLS
• *RST
none
none
*OPC
*OPC?
*OPC?
The OPeration Complete query *OPC? parses all program message units in the input
queue, sets the operation complete bit in the Standard Event Status register, and places
an ASCII '1' in the output queue, when the contents of the input queue have been pro-
cessed.
The following actions cancel the *OPC? query (and put the instrument into Operation
Complete, Command Idle State):
• Power-on
• the Device Clear Active State is asserted on the interface.
• *CLS
• *RST
none
1<END> is always returned.
*OPC? → 1<END>
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
IEEE-Common Commands