Signal Generation – The SOURce Subsystem
command:
syntax:
NO TE
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
[:SOURce[n]][:CHANnel[m]]:POWer:WAVelength?
[:SOURce[n]][:CHANnel[m]:POWer:WAVelength?
For compatibility reasons, WAVelength may be replaced with WAVE.
Returns the wavelength source for a dual-wavelength laser source.
none
LOW
UPP
BOTH
sour2:pow:wav? → LOW<END>
All dual-wavelength laser source modules and return loss modules with two internal
sources
[:SOURce[n]][:CHANnel[m]]:READout:DATA?
[:SOURce[n]][:CHANnel[m]:READout:DATA?
Returns the data as a binary stream from either a lambda logging operation or the maxi-
mum power the laser can produce at each wavelength.
LLOGging:
Returns a binary stream that contains each wavelength step of the lambda
logging operation, see "[:SOURce[n]][:CHANnel[m]]:WAVe-
length:SWEep:LLOGging" on page 111. Each binary block is an 8-byte long
double in Intel byte order.
PMAX:
Returns a binary stream that contains the maximum power the laser can
produce at each wavelength. Each binary block is a 8-byte long double (the
wavelength value) followed by a 4-byte long float (the power value). The
stream is in Intel byte order.
A binary stream in Intel byte order.
sour2:read:data? llog → the data as a binary stream
All tunable laser and DFB source modules
[:SOURce[n]][:CHANnel[m]]:READout:POINts?
[:SOURce[n]][:CHANnel[m]:READout:POINts?<wsp>LLOGging|PMAX
Returns the number of datapoints that the [:SOURce[n]][:CHANnel[m]]:READout:DATA?
command will return.
LLOGging:
Returns the number of wavelength steps for a lambda logging operation,
see "[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:LLOGging" on
page 111.
PMAX:
Returns the number of datapoints (each datapoint contains a value for
wavelength and power) the [:SOURce[n]][:CHANnel[m]]:READout:DATA?
PMAX command will return, number of datapoints depends on the calibra-
tion data for your module.
The number of datapoints as an integer value.
sour2:read:poin? pmax → 120<END>
All tunable laser and DFB source modules
Measurement Operations & Settings
The lower wavelength source
The upper wavelength source
Both wavelength sources