Agilent Technologies 8163A Programming Manual page 58

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Instrument Setup and Status
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
58
:STATusn:OPERation:CONDition?
:STATusn:OPERation:CONDition?
Returns the Operational Slot Status Condition Register of slot n.
none
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
Mnemonic
4-15
Not used
3
Slot n: Zeroing ongoing
2
Not used
1
Slot n: Coherence Control is switched on
0
Slot n: Laser is switched on
stat1:oper:cond? → +0<END>
:STATusn:OPERation:ENABle
:STATusn:OPERation:ENABle<wsp><value>
Sets the bits in the Operation Slot Status Enable Mask (OSSEM) for slot n that enable the
contents of the Operation Slot Status Event Register (OSSER) for slot n to affect the OS-
ESR.
Setting a bit in this register to 1 enables the corresponding bit in the OSSER for slot n to af-
fect bit n of the OSESR.
The bit value for the OSSEM as a 16-bit signed integer value (0 .. +32767)
none
stat:oper:enab 128
:STATusn:OPERation:ENABle?
:STATusn:OPERation:ENABle?
Returns the OSSEM of slot n
none
The bit value for the OSSEM as a 16-bit signed integer value (0 .. +32767)
stat:oper:enab? → +128<END>
:STATus:PRESet
:STATus:PRESet
Presets all bits in all the enable masks for both the OPERation and QUEStionable status
systems to 0, that is, OSSEM, QSSEM, OSESM, and QSESM.
none
none
stat:pres
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Status Reporting – The STATus Subsystem
Decimal Value
0
8
0
2
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a8166a

Table of Contents