Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
affects:
110
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl<wsp>
<value>|MIN|MAX|DEF[NS|US|MS|S]
Sets the dwell time.
The dwell time as a float value.
If you specify no units in your command, seconds are used as the default.
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of,
the minimum programmable value and the maximum programmable
value
none
wav:swe:dwel 500ms
All tunable laser modules
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl?
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:DWELl?[<wsp>MIN|MAX|DEF]
Returns the dwell time.
none
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of,
the minimum programmable value and the maximum programmable
value
The dwell time in seconds.
wav:swe:dwel?
All tunable laser modules
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Signal Generation – The SOURce Subsystem
+5.00000000E-001<END>