Status Reporting – The STATus Subsystem
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:STATus:QUEStionable:ENABle[:LEVel0]?
:STATus:QUEStionable:ENABle[:LEVel0]?
Returns the QSESM for the event register
none
The bit value for the QSEM as a 16-bit signed integer value (0 .. +32767)
stat:ques:enab? → +128<END>
:STATus:QUEStionable[:EVENt]:LEVel1?
:STATus:QUEStionable[:EVENt]:LEVel1?
Returns the Questionable Status Event Summary Register (QSESR) for slots 15 to 17 of the
Agilent 8166A Lightwave Multichannel System.
none
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤ 32767):
Bits
15-4
Not used
3
Slot 17 Summary
2
Slot 16 Summary
1
Slot 15 Summary
0
Not used
stat:ques:level1? → +0<END>
:STATus:QUEStionable:CONDition:LEVel1?
:STATus:QUEStionable:CONDition:LEVel1?
Returns the Questionable Status Condition Summary Register for slots 15 to 17 of the
Agilent 8166A Lightwave Multichannel System.
none
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤ 32767):
Bits
15-4
Not used
3
Slot 17 Summary
2
Slot 16 Summary
1
Slot 15 Summary
0
Not used
stat:ques:cond:level1? → +0<END>
Mnemonics
Agilent 8166A
Mnemonics
Agilent 8166A
Instrument Setup and Status
Decimal Value
0
8
4
2
0
Decimal Value
0
8
4
2
0
61