Agilent Technologies 8163A Programming Manual page 57

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Status Reporting – The STATus Subsystem
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:STATus:OPERation:ENABle:LEVel1
:STATus:OPERation:ENABle:LEVel1<wsp><value>
Sets the bits in the Operational Status Enable Summary Mask (OSESM) that enable the con-
tents of the OSESR for slots 15 - 17 of the Agilent 8166A Lightwave Measurement System to
affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR for slots 15 - 17
of the Agilent 8166A Lightwave Measurement System to affect bit 7 of the Status Byte.
The bit value for the OSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
none
stat:oper:enab:level1 128
:STATus:OPERation:ENABle:LEVel1?
:STATus:OPERation:ENABle:LEVel1?
Returns the OSESM for the OSESR for slots 15 - 17 of the Agilent 8166A Lightwave Measure-
ment System
none
The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
stat:oper:enab:level1? → +128<END>
:STATusn:OPERation[:EVENt]?
:STATusn:OPERation[:EVENt]?
Returns the Operational Slot Status Event Register (OSSER) of slot n.
none
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
Mnemonic
4-15
Not used
3
Slot n: Zeroing started
2
Not used
1
Slot n: Coherence Control has been switched on
0
Slot n: Laser has been switched on
stat1:oper? → +0<END>
Instrument Setup and Status
Decimal Value
0
8
0
2
1
57