Program Overall - Agilent Technologies E5061B Service Manual

Network analyzers
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Troubleshooting
Troubleshooting Using Diagnostic Test
Step 6. Double-click "Service" folder to open it.
Step 7. Click "DiagnosticTest. VBA" program file to select it from program menu.
Step 8. Click "Open" button to download the diagnostic test program.
Step 9. Press
Step 10. Press
dialog box (Main Menu) will appear as shown in Figure 2-14.
Figure 2-14
Diagnostic Test dialog box
Step 11. To exit the diagnostic test, click "Exit" button.

Program Overall

Instrument
Information
Test Conditions
Test Items
32
to select
Select Macro
Select Macro
to open the
Module1 main
The option and serial number for the E5061B, and test date can be
entered automatically.
The test conditions, "Temperature", "Humidity" and person doing the
test ("Tested by") can be entered.
The following test can be selected. To execute the test you select, click
"Execute" button.
DC-BUS Test
Internal Level Monitor Test
RF Output Level Range Test
RF Output Level Power SweepTest
Receiver Absolute Measurement Test
Receiver Compression Test (Option 3L5 only)
Receiver IF Ranging Test
Gain-Phase Absolute Measurement Test (option 3L5 only)
When each test is executed, the test dialog box is displayed. The
following screen is an example of the test dialog box.
function.
program file. "Diagnostic Test Program"
Module1 main
Chapter 2