Repair and Replacement Procedures

Removing and Replacing the A31, A32, A33, and A34 Receiver Modules and LO Reject Filter Board

Removing and Replacing the A31, A32, A33, and A34
Receiver Modules and LO Reject Filter Board
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-23
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
analyzer bottom side up and raise the receiver deck as shown. Refer to
Receiver Deck" on page 7-8
3. Identify the receiver module to be removed. To replace one of the receiver modules, the
LO reject filter board (item
4. Disconnect the flexible cables (item
modules.
5. Disconnect the ribbon cable (item
6. With a T-10 TORX driver, remove the eight mounting screws (item
reject filter board. Remove the board from the receiver motherboard.
7. With a T-10 TORX driver, remove the three mounting screws (item
appropriate receiver module.
8. Remove the receiver module from the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-50
for this procedure.
"Removing the Covers" on page
if necessary.
) must be removed from the receiver motherboard.
) from the LO reject filter board and the receiver
) from the LO reject filter board.
Table 7-2 on page
PNA Series Microwave Network Analyzers
7-6. Position the
"Raising the
) from the LO
) from the
7-66.