This chapter identifies the specifications, classifications, & system features for the Trilogy Ventilators.
3.1 E
NVIRONMENTAL
Temperature
Relative Humidity
Atmospheric Pressure
The operating range for the charging of the internal and detachable batteries is 10° C to 30° C.
The internal and detachable batteries will power the ventilator for the full operating range of 5°
C to 40° C.
Accuracies stated in this manual are based on specific environmental conditions. For stated
accuracy, the environmental conditions are: Temperature: 20-30° C, Humidity: 50% relative,
Altitude: nominally 380 meters
3.2 T
100 & T
RILOGY
Dimensions
Weight
3.3 T
O
RILOGY
2
Dimensions
Weight
1002735, V
. 05
ER

PECIFICATIONS

S
PECIFICATIONS
Operating
5° C to 40° C
15 to 95% (non-condensing)
110 kPa to 60 kPa
200 P
RILOGY
16.68 cm L x 28.45 cm W x 23.52 cm H
Approximately 5 kg (with the detachable battery installed)
& T
202 P
RILOGY
21.13 cm L x 28.45 cm W x 23.52 cm H
Approximately 6.1 kg (with the detachable battery installed)
T
S
RILOGY
ERVICE
, C
LASSIFICATIONS
S
HYSICAL
PECIFICATIONS
S
HYSICAL
PECIFICATIONS
& T
I
ECHNICAL
NFORMATION
, & S
Storage
-20° C to 60° C
15 to 95% (non-condensing)
N/A
YSTEM
P
3-1