15. Save it to the SD card again. In addition to the .exe file, there will now also be a .s file. This is the
file that the Trilogy needs to upgrade to the new software.
16. Remove the SD Card from the SD Card Reader and insert it into the Trilogy Device while in ther-
apy off mode.
Upgrading the Trilogy firmware will reset the device prescription and alarm settings to factory defaults.
If you are upgrading this device for use on the same patient, ensure you record all prescription and
alarm settings prior to upgrading the Trilogy device. Refer to the Trilogy Clinical Manual for details.
17. A prompt will then appear on your Trilogy User Interface asking you if you would like to upgrade
the Operating Software.
18. Select the YES button and follow the on-screen prompts to complete the Firmware installation.
1002735, V
. 05
ER
WARNING
T