Respironics Trilogy 100 Technical Information page 89

Table of Contents

Advertisement

A
P
LARM
RIORITY
High Expiratory
High
Pressure
Low Expiratory
High
Pressure
High Internal
High
Oxygen
P
6-8
AGE
D
ESCRIPTION OF
A
UDIBLE
It occurs when the
• • •
• •
delivered pressure
exceeds the target
patient pressure during
the expiratory phase by
5 pressure units. This
may be due to pinched
tubing or the patient
having a fast breath
rate. The alarm will
automatically terminate
when the delivered
pressure comes within
5 pressure units during
the expiratory phase.
It occurs when the
• • •
• •
delivered pressure is 5
pressure units or more
below the target patient
pressure during the
expiratory phase. The
alarm will automatically
terminate when the
delivered pressure
comes within 5
pressure units of the
target patient pressure
during the expiratory
phase.
It occurs when there is
• • •
• •
a leak in the internal air
delivery system that
allows oxygen to build
up inside the device.
The alarm is generated
when the internal
oxygen concentration
reaches 5% above
ambient levels.
T
S
& T
RILOGY
ERVICE
ECHNICAL
V
ISUAL
I
NDICATOR
A
B
LARM
UTTON AND
D
Red flashing
button; "High
Expiratory
Pressure"
message
Red flashing
button; "Low
Expiratory
Pressure"
message
Red flashing
button; "High
Internal Oxygen"
message
I
NFORMATION
(A
LARM
D
EVICE
A
CTION
)
ISPLAY
Operates
Operates
Continues
to operate
when
internal
oxygen
concentra
tion
reaches
5% above
ambient
levels.
1002735, V
ER
. 05

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trilogy 200Trilogy 02Trilogy 202