Respironics Trilogy 100 Technical Information page 322

Table of Contents

Advertisement

V
L
I
ERIFY THE
OW
NSPIRATORY
This procedure verifies that the Low Inspiratory Pressure alarm is working properly. It assumes that you have
attached the test lung, verified ventilator settings, and turned on ventilator power as described in the Initial
Setup section.
Change Ventilator Settings
Modify the ventilator settings and change the following values shown below.
Mode
Tidal Volume
Breath Rate
Inspiratory Time
Flow Pattern
PEEP
Sigh
Circuit Disconnect
Low Inspiratory Pressure
High Inspiratory Pressure
Apnea
All other alarms
Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The High Priority audible indicator sounds
A red light flashes on the Alarm Indicator/Audio Pause button
The Low Inspiratory Pressure alarm condition appears on the screen, highlighted in red
Modify Ventilator Alarm Settings
Modify the Low Inspiratory Pressure setting to match the one shown below.
Low Inspiratory Pressure
1002735, V
. 05
ER
P
A
RESSURE
LARM
SETTING
SETTING
T
S
& T
RILOGY
ERVICE
VALUE
CV
500 ml
12 BPM
1.0 seconds
Ramp
4 pressure units
Off
Off
40 pressure units
60 pressure units
Off
Off
VALUE
6 pressure units
I
ECHNICAL
NFORMATION
P
10-19
AGE