Respironics Trilogy 100 Technical Information page 328

Table of Contents

Advertisement

Options Menu
Modify the settings in the Options menu to match those shown below.
Menu Access
Detailed View
All other settings
Turn Device Power On
Press the Start/Stop button on the front of the ventilator. The system will begin operating using the defined
ventilation settings.
V
H
T
ERIFY THE
IGH
IDAL
This procedure verifies that the High Tidal Volume alarm is working properly. For Passive and Active Flow
circuits, this will verify the High Vte alarm. For Active PAP circuits, this will verify the High Vti alarm. It assumes
that you have attached the test lung, verified the ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
Do not use the "reset" button to manually reset the alarm. Instead, use the "Modify"
button to change ventilator settings. This note applies to all tests.
Change Alarm Ventilator Setting
Modify the High Tidal Volume alarm setting to match the one shown below.
High Vte/High Vti
Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The High Priority audible indicator sounds
A red light flashes on the Alarm Indicator/Audio Pause button
The High Tidal Volume alarm condition appears on the screen, highlighted in red
Modify Ventilator Alarm Settings
Modify the High Tidal Volume alarm setting to match the one shown below.
High Vte/High Vti
1002735, V
. 05
ER
SETTING
V
A
OLUME
LARM
SETTING
SETTING
T
S
RILOGY
ERVICE
VALUE
Full
On
Discretionary
NOTE
VALUE
50 ml
VALUE
500 ml
& T
I
ECHNICAL
NFORMATION
P
10-25
AGE