Respironics Trilogy 100 Technical Information page 95

Table of Contents

Advertisement

A
P
LARM
RIORITY
Low Inspiratory
High
Pressure
High Minute
High
Ventilation
P
6-14
AGE
D
ESCRIPTION OF
A
UDIBLE
• • •
• •
It is detected differently
for volume and
pressure therapy
modes.
For volume modes,
the alarm will sound if
the measured patient
pressure is less than
the Low Inspiratory
Pressure setting by the
clinician. The alarm
will automatically
terminate when the
peak pressure at the
end of the breath is
greater than or equal to
the Low Inspiratory
Pressure alarm setting.
For pressure modes,
the alarm occurs when
the delivered pressure
is 5 pressure units or
more below the target
patient pressure during
the inspiratory phase.
the alarm will
automatically terminate
when the delivered
pressure comes within
5 pressure units of the
target patient pressure
during the expiratory
phase.
• • •
• •
It occurs when the
patient's minute
ventilation is greater
than the High Minute
Ventilation alarm
setting. The alarm will
automatically terminate
when the calculated
minute ventilation is
less than the High
Minute Ventilation
alarm setting.
T
S
& T
RILOGY
ERVICE
ECHNICAL
V
ISUAL
I
NDICATOR
A
B
LARM
UTTON AND
D
Red flashing
button: "Low
Inspiratory
Pressure"
message
Red flashing
button; "High
Minute
Ventilation"
message
I
NFORMATION
(A
LARM
D
EVICE
A
CTION
)
ISPLAY
Operates
Operates
1002735, V
ER
. 05