E-84
Card Error
E-85
Card Error
E-86
Card Error
E-87
Card Error
E-88
Card Error
E-89
Card Error
E-93
Keypad Stuck
1002735, V
. 05
ER
Foreign object inserted in card
slot
Unformatted card
Card prematurely removed
Foreign object inserted in card
slot
Unformatted card
Card prematurely removed
Foreign object inserted in card
slot
Unformatted card
Card prematurely removed
Foreign object inserted in card
slot
Unformatted card
Card prematurely removed
Foreign object inserted in card
slot
Unformatted card
Card prematurely removed
Foreign object inserted in card
slot
Unformatted card
Card prematurely removed
Key is providing reading that it
has been held down for 2
minutes
T
S
& T
RILOGY
ERVICE
ECHNICAL
Use correct MMC/SD card
Re-insert card
Replace System/CPU PCA
Use correct MMC/SD card
Re-insert card
Replace System/CPU PCA
Use correct MMC/SD card
Re-insert card
Replace System/CPU PCA
Use correct MMC/SD card
Re-insert card
Replace System/CPU PCA
Use correct MMC/SD card
Re-insert card
Replace System/CPU PCA
Use correct MMC/SD card
Re-insert card
Replace System/CPU PCA
Check the Keypad for stuck keys
Replace the Front Panel PCA
Replace the System/CPU PCA