Respironics Trilogy 100 Technical Information page 119

Table of Contents

Advertisement

E-267
FCC > 150% of the Li-Ion design capacity
Corrupted Gas Gauge
E-269
Error indicates an open VBULK FET.
MOSFET connects the output of the AC/DC
supply to the input of the boost converter. Unit
will not run on AC power. Will use battery power
if available
E-270
Error indicates an open Q11, which controls the
discharge of the Li-Ion batteries post VbattLi.
When the Lithium Ion power is selected, this FET
is turned on. If Q11 fails open, there will be no Li-
Ion battery power to the system
E-272
Charger chip failure
FET failure
E-273
Unit not calibrated
Calibration table corrupted
E-274
Unit not calibrated
Calibration table corrupted
E-290
One or more "Ventilator Service Required" errors
are active in the system
E-302
Voltage reference on internal ADC failed
E-311
Unit wasn't calibrated or calibration table was
destroyed
P
6-38
AGE
T
S
& T
RILOGY
ERVICE
ECHNICAL
Cycle the Internal Li-Ion (discharge
battery so that capacity change is
> 50%, then rest for > 2 hours.
Charge to 100%, then rest for > 2
hours)
Replace Internal Li-Ion battery
This
Replace Power Management PCA
Replace Power Management PCA
Replace Power Management PCA
Recalibrate the device
Replace the System/CPU PCA
Recalibrate the device
Replace the System/CPU PCA
Examine the Significant Event Log
error log. The first column of data
contains the first error code that is
causing this error message.
Address according to this first error
code
Each time you address an error,
cycle the motor (OFF to ON to
OFF) and check the Significant
Event Log again until this alarm is
no longer sounded
Remove all power from the unit
and then reapply power, then turn
the motor ON to see if this
message is still reported
Replace System/CPU PCA
Recalibrate
Replace System/CPU PCA
I
NFORMATION
1002735, V
. 05
ER