Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The Medium Priority audible indicator sounds
A yellow light flashes on the Alarm Indicator/Audio Pause button
The High Inspiratory Pressure alarm condition appears on the screen, highlighted in yellow
Modify Ventilator Alarm Settings
Modify the High Inspiratory Pressure setting to match the one shown below.
High Inspiratory Pressure
Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The Medium Priority audible indicator has stopped sounding
The yellow light on the Alarm Indicator/Audio Pause button has stopped flashing
V
L
I
ERIFY THE
OW
NSPIRATORY
This procedure verifies that the Low Inspiratory Pressure alarm is working properly. It assumes that you have
attached the test lung, verified ventilator settings, and turned on ventilator power as described in the Initial
Setup section.
1002735, V
. 05
ER
SETTING
P
A
RESSURE
LARM
T
S
& T
RILOGY
ERVICE
VALUE
60 pressure units
I
ECHNICAL
NFORMATION
P
10-29
AGE