Respironics Trilogy 100 Technical Information page 194

Table of Contents

Advertisement

7. Remove the four screws securing the Power Management PCA Support to the Power Manage-
ment PCA.
Installation
1. Secure the Power Management PCA to the Power Management PCA Support by tightening the
screws to 8 in-lbs.
2. Align the locking tab with the slot in the Power Management PCA. While firmly pressing down
secure the Power Management PCA to the System PCA
3. Using the four screws, secure the Power Management PCA support to the System PCA support
by tightening the screws to 12 in-lbs.
4. Install the Metal Support Brackets. Refer to the Metal Support Bracket Replacement Section for
more details.
5. Install the PCA Inverter. Refer to the PCA Inverter Replacement Section for more details.
6. Install the Front Panel PCA. Refer to the Front Panel PCA Replacement Section for more details.
7. Install the Keypad/Front Enclosure. Refer to the Keypad/Front Enclosure Replacement Section for
more details.
1002735, V
. 05
ER
F
8-30: PCB S
IGURE
T
S
& T
RILOGY
ERVICE
ECHNICAL
S
UPPORT
CREWS
I
NFORMATION
P
8-33
AGE

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trilogy 200Trilogy 02Trilogy 202

Table of Contents