Respironics Trilogy 100 Technical Information page 91

Table of Contents

Advertisement

A
P
LARM
RIORITY
Low Oxygen
High
Inlet Pressure
(Trilogy O
&
2
Trilogy 202
Only)
Circuit
High
Disconnect
Apnea
High
P
6-10
AGE
D
ESCRIPTION OF
A
UDIBLE
• • •
• •
It occurs when the
pressure of the oxygen
from the oxygen source
measures less than 40
psi. This could be
caused by the oxygen
source being
disconnected from the
device, an occlusion in
the tubing from the
oxygen source to the
device, or a problem
with the output of the
oxygen source.
It occurs when the
• • •
• •
breathing circuit is
disconnected or has a
large leak. The alarm
will automatically
terminate when the
leak is fixed for 6
seconds and breaths
are delivered again.
It occurs when the
• • •
• •
patient has not
triggered a breath
within the time
specified in the apnea
alarm setting. The
alarm will automatically
terminate when two
consecutive patient
breaths are detected
that meet the apnea
alarm time setting.
T
S
& T
RILOGY
ERVICE
ECHNICAL
V
ISUAL
I
NDICATOR
A
B
LARM
UTTON AND
D
Red flashing
button; "Low
Oxygen Inlet
Pressure"
message
Red flashing
button; "Circuit
Disconnect"
message
Red flashing
button; "Apnea"
message
I
NFORMATION
(A
LARM
D
EVICE
A
CTION
)
ISPLAY
Operates
Operates
Operates
1002735, V
ER
. 05

Advertisement

Table of Contents
loading

This manual is also suitable for:

Trilogy 200Trilogy 02Trilogy 202