In the next steps, you will be asked to remove the 240 VAC and connect a meter into the
DC cable to measure the DC Current. Please follow the steps below carefully otherwise
an E280 error will be logged in the Error Log and will result in the test failing.
11. Remove 240 VAC, observe the AC Disconnected info message on the Trilogy Display.
12. After the AC Disconnected message is received, press the reset button on Trilogy Device.
Observe the Lead Acid icon is green and a black box is around the green Lead Acid icon.
13. Break the connection in the DC cable, and WAIT for the info message of Lead Acid Disconnected
on the Trilogy Display.
14. Connect the Meter in series with the DC Cable. Press reset on the Trilogy Device and observe that
the Lead Acid icon is green, and a black box is around the green Lead Acid icon.
1002735, V