Respironics Trilogy 100 Technical Information page 116

Table of Contents

Advertisement

E-134
Control flow sensor is reading maximum or
minimum counts for an extended period of time
E-135
Monitor flow sensor is reading maximum or
minimum counts for an extended period of time
E-136
Air stream Temperature 1 sensor is reading
maximum or minimum counts for an extended
period of time
E-137
dP Pressure Temperature sensor is reading
maximum counts for an extended period of time
E-138
Barometric Pressure sensor is reading a count
outside of its valid range
E-139
Software Error
E-143
Control or Monitoring Pressure sensor has
drifted too far away from the other
1002735, V
. 05
ER
T
S
& T
RILOGY
ERVICE
ECHNICAL
Check tubing
Check manifold
Check the Sensor Board Cable
Replace the Sensor PCA
Replace the System/CPU PCA
Check tubing
Check manifold
Check the Sensor Board Cable
Replace the Sensor PCA
Replace the System/CPU PCA
Check Airstream Temp Sensor
connections
Replace Sensor
Replace Sensor Board
Replace System/CPU PCA
Check dP Pressure Temp Sensor
Replace Sensor
Replace Sensor Board PCA
Replace System/CPU PCA
Replace the Sensor PCA
Replace the System/CPU PCA
Recalibrate the device
Replace the System/CPU PCA
Turn off Blower and allow 30
seconds for drift processing to
complete
Replace the Sensor PCA
I
NFORMATION
P
6-35
AGE