Chapter 8: Specifications

Default Settings

S/T mode
IPAP = 12
cmH
O
2
EPAP = 4
cmH
O
2
Rise-Time = 2

Settings: CPAP Mode

CPAP
Continuous
Positive Airway
Pressure
Ramp Time
Period over
which the
ventilator
increases
inspiratory
pressure from
Ramp Start
setting to CPAP
setting.
Ramp Start
Initial
inspiratory
pressure
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
RR = 4/min
I-Time = 1 sec
Ramp Time = 0 min
Ramp Start = 4 cmH
Range: 4-20 cmH
Resolution: 1 cmH
Dynamic accuracy: ±5 cmH
Range: 0-45 min.
Resolution: 5 min.
Accuracy: ±10% of setting
Range: 4 cmH
O to CPAP setting (4 hPa to CPAP setting)
2
Resolution: 1 cmH
O
2
O (4-20 hPa)
2
O (1 hPa)
2
O (5 hPa)
2
O (1 hPa)
2