Respironics Trilogy 100 Technical Information page 122

Table of Contents

Advertisement

E-199
Detachable Battery harness failure
Power Management Board Failure
Circuitry failure
E-200
Circuitry failure
E-201
Power Management Board Failure
Circuitry failure
E-202
Circuitry failure
E-204
U3 failure
CPLD failure
E-271
Error indicates an open Q13, Q14 in the Lead Acid
battery discharge path on the Power Management
PCA. When Pb Acid is selected by the hardware
or software, VBATTSENSE would be low. If AC is
disconnected, there would be no battery backup
from the Li-Ion battery
E-289
One or more "Ventilator Service Recommended"
errors are active in the system
E-304
O
sensor failed
2
E-345
Internal O
E-354
Unit not calibrated
O
Sensor Calibration Table corrupted
2
1002735, V
. 05
ER
sensor at end of life or watchdog failure
2
T
S
& T
RILOGY
ERVICE
Replace
harness
Replace
Management PCA
Replace the System/CPU PCA
Replace the System/CPU PCA
Replace
Management PCA
Replace the System/CPU PCA
Replace the System/CPU PCA
Replace Interface PCA
Replace System/CPU PCA
Replace Power Management
PCA
Examine the Significant Event
Log error log. The first column
of data contains the first error
code that is causing this error
message
Address according to this first
error code
Each time you address an error,
cycle the motor (OFF to ON to
OFF) and check the Significant
Event Log again until this alarm
is no longer sounded
Remove all power from the unit
and then reapply power, then
turn the motor ON to see if this
message is still reported
Replace Interface PCA
Replace OBM PCA
Recalibrate
Replace System/CPU PCA
I
ECHNICAL
NFORMATION
Detachable
Battery
the
Power
the
Power
P
6-41
AGE