This chapter illustrates the replaceable components for the Trilogy Ventilators. Procedures for replacing the
components are also provided in this chapter.
Refer to the Testing Section for required testing after component
replacement.
Refer to the Repair Kits Section for proper repair kit identification.
To prevent electrical shock, disconnect the electrical supply before attempting to make
any repairs to the Trilogy Ventilators.
Electronic components used in this device are subject to damage from static electricity.
Repairs made to this device must be performed only in an antistatic, ESD-protected
environment.
During all repair and replacement procedures, ensure that any connections that are
broken during the process (fittings, tubing, and hoses) are reconnected securely.
When using a leak detector, be careful that it does not come in contact with any
electrical components.
Using the Trilogy Toolbox Application, place the unit in "Ship Mode" prior to any
disassembly of the unit. Otherwise, ensure that the Internal Battery connection is the
first to be disconnected and last to be reconnected during re-assembly.