5. Using a standoff, secure the Mixing Element to the PCA be securing the screw to 6 in-lbs.
6. Reconnect the tubing to the Flow Element Tubes and ensure it is connected properly to the PCA
flow sensors.
7. Ensuring that the two small o-rings are in place reconnect the Pressure Sensor Clip to the PCA.
8. Install the Oxygen Blender Housing. Refer to Oxygen Blender Housing Removal Section for more
details.
9. Install the Filter/Whisper Cap. Refer to Oxygen Blender Filter/Whisper Cap Replacement Section
for more details.