Performance Verification - Respironics Trilogy 100 Technical Information

Table of Contents

Advertisement

I
TEM
Porting Blocks
Detachable Battery (if present)
Handle
Rubber Feet
Power Cord
External Tubing
Cables and Internal Tubing
Cleaning
Perform Field Service Application
(FSA) Post Test Procedure or
Performance Verification (PV) Tool
a
If the unit fails the FSA Post Test Procedure refer to the Troubleshooting Chapter of this service
manual to determine the corrective action that needs taken. Once the corrective action has been
performed run the complete FSA.
b
If the unit fails the PV tool, return the device to an authorized Service Facility for troubleshooting and/or
repair. Once the corrective action has been performed run the complete FSA.
P
7-6
AGE
Check O-Rings around the porting blocks for cracks, wear, or
incorrect fit. Lubricate or replace the porting block as
necessary.
Verify the latch and guide mechanism is intact and functional.
Compare the capacity LEDs on the Detachable Battery Pack to
the Detachable Battery Icon on the User Interface. The Charge
levels should be within 1 level of each other.
Verify the Handle is free of defects. Check O-Rings around the
handle for cracks, wear, or incorrect fit. Lubricate or replace as
necessary.
Verify that all of the Rubber Feet are present.
Inspect the power Cord for any cracks/cuts.
Clean or replace all external tubing as necessary.
Remove the Bottom Enclosure and verify the integrity of cable
connectors and internal tubing.
Remove the Bottom Enclosure and vacuum out the bottom of
the unit using an Anti-static vacuum.
Preform the Field Service Application Post Test Procedure
the Performance Verification Tool