8.3.6 C
/I
APACITOR
NTERNAL
Removal
1. Remove the three screws securing the Capacitor/Internal Battery Retainer to the Bottom Enclo-
sure.
F
IGURE
2. Cut the two tie wraps securing wires to the Capacitor/Internal Battery Retainer.
3. Remove the Capacitor and/or Internal Battery.
Installation
1. Place the Capacitor into its slot in the Bottom Enclosure.
2. Route the Internal Battery cable through the slot in the Internal Battery/Capacitor retainer.
3. Secure the Capacitor and/or Internal Battery to the Bottom Enclosure using the Capacitor/Battery
Retainer. Tighten the three screws to 8 in-lbs.
4. Using a tie wrap, secure the Capacitor wires to the Capacitor/Internal Battery Retainer.