Respironics Trilogy 100 Technical Information page 84

Table of Contents

Advertisement

6.2.3 A
M
LARM
ESSAGES
NOTE
An alarm message will also display in the Menu Banner if a menu is active when an alarm occurs.
When the ventilator detects an alarm, the Alarms and Messages Screen is displayed showing a description of
the alarm condition. When an alarm message appears, it will be highlighted in red if it is a high priority alarm or
in yellow if it is a medium or low priority alarm. (The highlight color matches the alarm LED color on the Alarm
Indicator/Audio Pause button.) If an alarm is manually reset by the user, the Alarms and Messages screen is
removed and the Monitoring Screen is re-displayed. If the alarm self-cancels, the Alarms and Messages
screen remains displayed, but the highlight from the active alarm is removed, the LED is unlit, and the audible
alarm stops. The screen below is an example of a possible alarm message.
If a menu is displayed on the screen when an alarm occurs, the description of the newly generated alarm is
displayed in the menu banner area. This is done so that the modification to the current parameter can be
completed before addressing the alarm condition in case the modification affects the alarm condition. The
screen below is an example of an alarm message displayed in the menu banner.
The Alarms and Messages Screen will automatically display in place of the Monitor screen when exiting from
the menu system using the Exit soft key when an alarm is displayed in the menu banner. If an alarm is
manually reset by the user or self-cancels, the menu banner on-screen before the alarm occurred will
reappear.