Respironics Trilogy 100 Technical Information page 318

Table of Contents

Advertisement

Restore Ventilator Settings
Modify the ventilator settings and change the following value shown below.
High Vte/High Vti
V
L
T
V
ERIFY THE
OW
IDAL
This procedure verifies that the Low Tidal Volume alarm is working properly. For Passive and Active Flow
circuits, this will verify the Low Vte alarm. For Active PAP circuits, this will verify the Low Vti alarm. It assumes
that you have attached the test lung, verified the ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
Change Alarm Ventilator Setting
Modify the Low Tidal Volume alarm setting to match the one shown below.
Low Vte/Low Vti
Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The High Priority audible indicator sounds
A red light flashes on the Alarm Indicator/Audio Pause button
The Low Tidal Volume alarm condition appears on the screen, highlighted in red
Modify Ventilator Alarm Settings
Modify the Low Tidal Volume alarm setting to match the one shown below.
Low Vte/Low Vti
Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The High Priority audible indicator has stopped sounding
The red light on the Alarm Indicator/Audio Pause button has stopped flashing
Restore Ventilator Settings
Modify the ventilator settings and change the following value shown below.
Low Vte/Low Vti
1002735, V
. 05
ER
SETTING
A
OLUME
LARM
SETTING
SETTING
SETTING
T
S
RILOGY
ERVICE
VALUE
Off
VALUE
500 ml
VALUE
50 ml
VALUE
Off
& T
I
ECHNICAL
NFORMATION
P
10-15
AGE