7.2.1 T
R
RILOGY
ECOMMENDED
T
/ H
IME
OURS OF
Prior to initial use on a patient
Prior to long term storage after initial
use
While in storage after initial
connection to AC Power, every 3
months
If in use, every two weeks
If in use, every 6 months
Every 10000 hours or 24 Months,
whichever comes first
Every 17,500 hours
P
7-4
AGE
M
AINTENANCE
S
ERVICE
•
Charge Internal and Detachable Batteries to 100%
capacity by plugging ventilator into an AC power source
for up to 8 hours
•
Perform System Checkout Procedure per Trilogy Clinical
Manual
•
Charge both Internal and Detachable batteries to 100%
capacity prior to storage.
•
Recharge internal and detachable batteries to 100%
capacity by plugging ventilator into an AC power source.
Batteries should recharge in 8 hours or less.
•
Inspect and clean air inlet filter
•
Replace air inlet filter
•
Inspect enclosures and external connections for damage
and contact Philips Respironics service if necessary
•
Inspect power cord for damage and replace if necessary.
•
Refer to appropriate section below.
•
Refer to appropriate section below.