Table of Contents

Advertisement

7.2.1 T
R
RILOGY
ECOMMENDED
T
/ H
IME
OURS OF
Prior to initial use on a patient
Prior to long term storage after initial
use
While in storage after initial
connection to AC Power, every 3
months
If in use, every two weeks
If in use, every 6 months
Every 10000 hours or 24 Months,
whichever comes first
Every 17,500 hours
P
7-4
AGE
M
AINTENANCE
S
ERVICE
Charge Internal and Detachable Batteries to 100%
capacity by plugging ventilator into an AC power source
for up to 8 hours
Perform System Checkout Procedure per Trilogy Clinical
Manual
Charge both Internal and Detachable batteries to 100%
capacity prior to storage.
Recharge internal and detachable batteries to 100%
capacity by plugging ventilator into an AC power source.
Batteries should recharge in 8 hours or less.
Inspect and clean air inlet filter
Replace air inlet filter
Inspect enclosures and external connections for damage
and contact Philips Respironics service if necessary
Inspect power cord for damage and replace if necessary.
Refer to appropriate section below.
Refer to appropriate section below.
T
S
& T
RILOGY
ERVICE
S
CHEDULE
R
M
ECOMMENDED
I
ECHNICAL
NFORMATION
AINTENANCE
1002735, V
. 05
ER

Advertisement