Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The High Priority audible indicator has stopped sounding
The red light on the Alarm Indicator/Audio Pause button has stopped flashing
10.3.5 B
F
ATTERY
UNCTION
Make sure the batteries are functioning properly and fully charged before patient use.
V
D
ERIFY THE
ETACHABLE AND
1. Connect AC Power to the device and verify that the green AC LED on the front panel is lit.
2. Verify that the detachable battery is properly installed.
3. Turn the device on and verify that both the detachable and internal battery symbols appear on the
display. Verify that if either battery is less than fully charged, the charge symbol will display on the
respective battery.
4. Disconnect the AC Power source from the device.
Verify that the AC Power Disconnected alarm message appears on the display and the green
AC LED is not lit. Press Reset.
Verify that the detachable battery symbol shows the level of charge noted in the previous step
and that the device continues to operate.
Verify that the detachable battery symbol has a black box around it to indicate that it is in use.
5. Disconnect the detachable battery pack from the device.
Verify that the Detach Batt Disconnected alarm message appears on the display. Press Reset.
Verify that the internal battery symbol shows the same level of charge as noted in Step C and
the device continues to operate.
Verify that the internal battery symbol has a black box around it to indicate that it is in use.
6. Reconnect the Detachable Battery and AC Power source.
V
E
B
ERIFY THE
XTERNAL
1. Connect AC Power to the device and verify that the green AC LED is lit.
2. Connect the external battery cable to the external battery and to the ventilator.
3. Verify that the external battery symbol is shown on the display and some level of charge is present.
4. Disconnect the AC Power source from the device.
Verify that the AC Power Disconnected alarm message appears on the display and the green
AC LED is not lit. Press Reset.
Verify that the external battery symbol shows the level of charge as noted in the previous step
and the device continues to operate.
Verify that the external battery symbol has a black box around it to indicate that it is in use.
5. Reconnect the AC Power source.
A
E
L
LARM AND
VENT
OG
1. In the Setup Menu, select Alarm Log.
a. Press Clear to clear the log file.
b. Press Yes to confirm.
P
10-20
AGE
V
ERIFICATION
I
(L
NTERNAL
ITHIUM
F
(I
A
ATTERY
UNCTION
F
VAILABLE
C
-U
LEAN
P
T
S
RILOGY
ERVICE
-I
) B
F
ON
ATTERIES
UNCTION
)
& T
I
ECHNICAL
NFORMATION