Respironics Trilogy 100 Technical Information page 101

Table of Contents

Advertisement

A
P
LARM
RIORITY
Detachable
Info
Battery
Disconnected
Start On
Info
Battery
Card Error
Info
P
6-20
AGE
D
ESCRIPTION OF
A
UDIBLE
This info message
occurs when the
detachable battery
power source is lost
and the device has
switched to an
alternate power source
(the internal battery if
not other source is
available). If
detachable battery
power returns, the
ventilator will beep, but
no message will appear
on the display.
This info message
indicates that the
ventilator has started
on battery power and
no AC power is
available. The device
operator should verify
that this is what is
wanted.
This info message
occurs when an
unusable SD card is
inserted into the
ventilator. The device
continues to operate
but data cannot be
logged onto the SD
card.
T
S
& T
RILOGY
ERVICE
ECHNICAL
V
ISUAL
I
NDICATOR
A
B
LARM
UTTON AND
D
"Detachable Batt
Disconnected"
message, and a
box appears
around battery in
use.
"Start On Battery"
message
appears.
"Card Error"
message appears
I
NFORMATION
(A
LARM
D
EVICE
A
CTION
)
ISPLAY
Switches
to
alternate
power
source
Operates
Operates
1002735, V
ER
. 05