19. Slide the Bottom Enclosure away from the Front and Rear Enclosures. Proceed to the Bottom
Enclosure Assembly Component Removal/Installation section for instructions on how to remove/
install the Bottom Enclosure Assembly components.
20. Remove the four screws securing the Front and Rear Enclosures together.
21. While holding the enclosures together place the Trilogy on its top so the opening is facing up.
22. Gently slide the Front and Rear Enclosures apart.
23. Remove the Sensor Cable from J4 on the Sensor Board.
24. Remove the Blower Cable from J5 on the System Board.
25. Remove the Blower Fan Cable from J17 on the System Board.
26. Remove the tape securing the Exhalation Control wires to the metal support.
27. Remove the Exhalation Control Module wires from J10 on the System Board.
28. Remove the Exhaust Fan wires from J1 on the System Board.
29. Remove the Detachable Battery Cable from J1 and J3 on the Power Management Board.
30. Cut the two tie wraps securing the Blower wires to the plastic support bracket.
31. Slide the Front and Rear Enclosures Apart.
32. Remove the Top Plate from the device.
33. The Front and Rear Enclosure are now disconnected. Proceed to the Front or Rear Enclosure
Assembly Component Removal/Installation sections for instructions on how to remove/install the
Assembly components. Also, Refer to the Oxygen Blending Module Components Section for
instructions on how to remove/install the Oxygen Blending Module components.