If a Ventilator Inoperative alarm occurs, the entire display screen turns red and the Ventilator inoperative
message appears on-screen, as shown below.
To turn the ventilator off from a Ventilator Inoperative condition, use the normal power off sequence. When the
Start/Stop button is selected, the following screen will display.
Select the Right button (Yes) to turn the ventilator off and stop the audible alarm. Selecting the Left button (No)
will return the screen to the Ventilator Inoperative Alarm Screen without silencing the audible alarm.
6.2.4 R
A
EMOTE
LARM
When the ventilator detects an alarm condition, if you are using a remote alarm system, a signal is sent from
the ventilator to activate the remote alarm.
6.3 A
P
UDIO
AUSE AND
This section describes the Audio Pause and Alarm Reset features.
6.3.1 A
P
UDIO
AUSE
When an alarm occurs, you can temporarily silence the audible indicator by pressing the Alarm Indicator/Audio
Pause button. The alarm is silenced for 60 seconds and then will sound again if the cause of the alarm has not
been corrected. Each time the Alarm Indicator/Audio Pause button is pressed, the alarm silence period resets
to one minute.