10.9.1 T
M
OOLBOX
ENU
L
M
OG
ENU
The Log Menu allows you to select what you would like to do with the information displayed on the Trilogy
Toolbox screen. Once the Trilogy Toolbox is exited, the information displayed on the screen is erased.
a. Clear – Clear the information currently displayed on the Toolbox Screen.
b. Print – Print the information currently displayed on the Toolbox Screen.
c. Save – Save the information currently displayed on the Toolbox Screen to a ".txt" file in a
folder of your choice for archiving and/or viewing at a later time.
d. Print As Report – Print the information currently displayed on the Toolbox Screen in a more
formal format which can include comments and Date – Time information.
S
M
ETUP
ENU
The Setup Menu provides tools for manipulating the conditions of the Unit Under Test (UUT). These tools can
aid in the troubleshooting of the Trilogy Device by turning on/off certain components as well as changing
certain settings.
a. TV Cal Mode On/Off – In certain situations the UUT can become stuck in the on or off condi-
tion and the need arises to change that condition. This option will allow that change; from an
on to an off state and vice versa.
b. Real Time Clock – Sets the Real Time Clock (RTC) of the UUT to the time of the PC perform-
ing the testing.