Respironics REMstar Plus with C-Flex User Manual

Respironics REMstar Plus with C-Flex User Manual

Continuous positive airway pressure

Advertisement

Symbol Key ................................................................................... 2
Intended Use & Contraindications ............................................. 3
Specifications ............................................................................... 4
Warnings ....................................................................................... 5
Cautions ........................................................................................ 6
Unpacking the System ................................................................. 6
System Features .......................................................................... 7
First Time Setup ........................................................................... 8
Installing the Filter .................................................................. 9
Device Settings .................................................................... 10
Assembling the Circuit ......................................................... 15
Accessories ................................................................................ 16
Adding a Humidifier .............................................................. 16
Adding Oxygen to the System .............................................. 16
Using DC Power with the System ........................................ 16
Routine Use ................................................................................ 17
Connecting the Circuit .......................................................... 17
Adjusting the Circuit ............................................................. 17
Ramp Button ........................................................................ 18
Turning the System Off ........................................................ 18
Helpful Hints ......................................................................... 18
Routine Maintenance ................................................................. 19
Cleaning the Filter ................................................................ 19
Cleaning the System ............................................................ 19
Reordering ........................................................................... 20
Service ................................................................................. 20
Traveling with the System ......................................................... 20
Troubleshooting ......................................................................... 21
EMC Requirements .................................................................... 23
Warranty ...................................................................................... 26

Table of Contents

REMstar Plus with C-Flex User Manual
1

Advertisement

Table of Contents
loading

Summary of Contents for Respironics REMstar Plus with C-Flex

  • Page 1: Table Of Contents

    Routine Maintenance ..............19 Cleaning the Filter ..............19 Cleaning the System ............19 Reordering ................20 Service ................. 20 Traveling with the System ............20 Troubleshooting ................. 21 EMC Requirements ..............23 Warranty ..................26 REMstar Plus with C-Flex User Manual...
  • Page 2: Symbol Key

    Symbols 0123 REMstar Plus with C-Flex CPAP systems are the subject of the following U.S. patents: #5239995 and #6105575. Other patents pending. REMstar, Whisper Swivel and Encore are registered trademarks of Respironics, Inc. ©Respironics, Inc. 2003 REMstar Plus with C-Flex User Manual...
  • Page 3: Intended Use & Contraindications

    Positive Airway Pressure) device designed for the treatment of adult Obstruc- tive Sleep Apnea (OSA) only. The REMstar Plus with C-Flex is to be used only on the instruction of a licensed health care professional. Your home care provider will make the correct pressure settings according to your health care professional’s...
  • Page 4: Specifications

    O @ 500 ml with BPM set to 10, 15, & 20 BPM performed at 23° C (±2° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals. ±(0.15 + 4% of actual reading) Pressure Display Accy. REMstar Plus with C-Flex User Manual...
  • Page 5: Warnings

    • Contact your health care professional if symptoms of sleep apnea recur. • Using the REMstar Plus with C-Flex at an incorrect elevation setting could result in airflow pressures higher than the prescribed setting. Always verify the elevation setting when traveling or relocating.
  • Page 6: Unpacking The System

    Warnings & Cautions/Unpacking the System CAUTION! • Tobacco smoke may cause tar build-up within the REMstar Plus with C-Flex that may result in the REMstar Plus with C-Flex malfunctioning. Indicates the possibility of damage • The use of accessories other than those specified, with the exception of to the device.
  • Page 7: System Features

    Medical Product Note ..For ease at airport security stations, there is a note on the bottom of the REMstar Plus with C-Flex stating that it is medical equipment. It may help if you also take this manual with you when you travel.
  • Page 8: First Time Setup

    Respironics or your health care provider. CAUTION! If the REMstar Plus with C-Flex has been exposed to either very hot or very cold temperatures, allow it to adjust to room temperature (approximately 2 hours) before beginning setup.
  • Page 9: Installing The Filter

    Inspect the power cord often for any signs of damage. Replace a damaged cord immediately. WARNING! The REMstar Plus with C-Flex is activated and ready for use when the power cord is connected. The pressure start/stop button turns the blower on/off.
  • Page 10: Device Settings

    First Time Setup (cont.) Device Once the power cord is plugged in, REMstar Plus with C-Flex’s system version number, the total hours of use, and then the total therapy hours Settings will appear. Display Screen: All device settings will appear here.
  • Page 11 IMPORTANT! airflow pressure, if prescribed, and then will gradually increase it. WARNING! DO NOT use the REMstar Plus with C-Flex if the display is erratic. Contact your home care provider for further instructions. IMPORTANT! If at any time you wish to exit the setup menu, press the pressure on/off button and the display will go back to the main screen.
  • Page 12 1 = on 0 = off To change the setting, press the ramp or humidifier button until the correct setting appears. Press the right user button to access the next setting. REMstar Plus with C-Flex User Manual...
  • Page 13 Press the right user button to access the next setting. WARNING! Using the REMstar Plus with C-Flex at an incorrect elevation setting could result in airflow pressures higher than the prescribed setting. Always verify the elevation setting when traveling or relocating.
  • Page 14 The settings are complete. Press the left user button to access the previous setting, or press the pressure start/stop button to exit the settings menu. Or, if you want to return to step a, press the right user button. REMstar Plus with C-Flex User Manual...
  • Page 15: Assembling The Circuit

    REMstar Plus with C-Flex flushes the exhaled air out through the exhalation port. When the REMstar Plus with C-Flex is turned off, enough fresh air will not be provided through the mask, and exhaled air may be rebreathed.
  • Page 16: Accessories

    Turn the oxygen off before turning the REMstar Plus with C-Flex off. Explanation of Warning: When the REMstar Plus with C-Flex is turned off, but the oxygen flow is still turned on, oxygen may accumulate within the REMstar Plus with C-Flex enclosure and create a fire risk.
  • Page 17: Routine Use

    If the airflow does not start within four breaths, press the pressure start/stop button on the top of the REMstar Plus with C-Flex. When operating the system with some mask types or some circuit configurations, the airflow may NOT automatically start.
  • Page 18: Ramp Button

    • Make sure that bedding, curtains, or other items are not blocking the filter or vents of the REMstar Plus with C-Flex. Air must flow freely around the REMstar Plus with C-Flex for the system to work properly.
  • Page 19: Routine Maintenance

    Gently wash the flexible tubing in a solution of warm water and a mild detergent. Rinse the tubing thoroughly and air dry. 2. Wipe the outside of the REMstar Plus with C-Flex with a cloth slightly dampened with water and a mild detergent. Let the REMstar Plus with C-Flex dry before plugging in the power cord.
  • Page 20: Reordering

    WARNING! If you notice any unexplained changes in the performance of the REMstar Plus with C-Flex, if it is making unusual or harsh sounds, if it has been dropped or mishandled, if the enclosure is broken, or if water has entered the enclosure, discontinue use.
  • Page 21: Troubleshooting

    Troubleshooting The table below lists common problems you may have with the REMstar Plus with C-Flex system and possible solutions to those problems. If none of the corrective actions solve the problem, contact your home care provider. Solution(s) Problem Press any button on the REMstar Plus with C-Flex to silence the The “Attention”...
  • Page 22 Problem (cont.) Solution(s) (cont.) The REMstar Plus with C-Flex will not Make sure that the REMstar Plus with C-Flex is plugged into a turn on. working outlet. If you are using DC power accessories with a battery, the battery may need to be recharged or replaced or the DC Cord’s fuse may need to be replaced.
  • Page 23: Emc Requirements

    ± ± ± ± i l a t s r . l a ± ± ± ± i l a l a i l a i . l a ± ± REMstar Plus with C-Flex User Manual...
  • Page 24 . l e REMstar Plus with C-Flex User Manual...
  • Page 25 EMC Requirements (cont.) REMstar Plus with C-Flex User Manual...
  • Page 26: Warranty

    (2) years from the date of sale by Respironics, Inc. to the dealer. If the product fails to perform in accordance with the product specifications, Respironics, Inc.
  • Page 27 If you have any questions concerning the system, contact: • Home Care Company: Telephone Number: • Health Care Professional: Telephone Number: • Respironics, Inc. 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668-8550 USA Customer Service Telephone Number: 1-800-345-6443 (USA or Canada) or 1-724-387-4000...

Table of Contents