This chapter details the test procedures that should be performed by the clinician prior to connecting the device
to the patient. Test the Active PAP, Active Flow, and Passive circuit types if you want to do a complete checkout
on the device. The tests should be performed as described in order to verify proper operation of the device.
Some of the procedures in this chapter require you to change settings on the device.
10.5.1 T
R
OOLS
EQUIRED
Universal Porting Block
Active PAP Exhalation Device
Active Flow Exhalation Device with Flow Sensor
Trilogy Universal Active PAP Tube Adapter
Whisper Swivel II
Test Lung
High Pressure O
Oxygen Monitor
Small Flat Head Screwdriver
10.5.2 V
I
ISUAL
NSPECTION
1. Verify that the enclosure is not broken and that all applicable screws are in place.
2. Verify that the device handle, SD Card door, and detachable battery are secure and in good work-
ing order.
3. Verify that the rubber feet are on the bottom of the device.
10.5.3 I
S
NITIAL
ETUP
1. Connect the power cord to the device and then to an AC outlet.
2. Attach the test lung to the patient connection end of the desired circuit (Active PAP, Active Flow, or
Passive).
3. Connect the device to a suitable high pressure O
4. Connect and set up an external O
5. Follow the instructions in Chapter 5 to access the Setup Screen.
1002735, V
. 05
ER
& T
202 C
RILOGY
Hose
2
2
T
S
RILOGY
ERVICE
P
HECKOUT
ROCEDURE
source.
2
monitor per the manufacturer's instruction manual.
& T
I
ECHNICAL
NFORMATION
P
10-23
AGE