I
TEM
Connectors
Porting Blocks
Detachable Battery (if present)
Handle
Rubber Feet
Power Cord
External Tubing
Cables and Internal Tubing
Cleaning
Perform the Complete Field Service
(FSA) Application (Pre and Post
Test)*
* If the unit fails the Field Service Application Pre and/or Post Test refer to the Troubleshooting section
of this service manual to determine the corrective action that needs taken. Once the corrective action
has been performed run the complete FSA.
P
7-8
AGE
Verify the integrity of connectors on the rear of the unit,
including the AC cord retainer.
Check O-Rings around the porting blocks for cracks, wear, or
incorrect fit. Lubricate or replace the porting block as
necessary.
Verify the latch and guide mechanism is intact and functional.
Compare the capacity LEDs on the Detachable Battery Pack
to the Detachable Battery Icon on the User Interface. The
Charge levels should be within 1 level of each other.
Verify the Handle is free of defects. Check O-Rings around
the handle for cracks, wear, or incorrect fit. Lubricate or
replace as necessary.
Verify that all of the Rubber Feet are present.
Inspect the power Cord for any cracks/cuts.
Clean or replace all external tubing as necessary.
Remove the Bottom Enclosure and verify the integrity of cable
connectors and internal tubing.
Remove the Bottom Enclosure and vacuum out the bottom of
the unit using an Anti-static vacuum.
Once the Blower Assembly has been replaced preform the
Field Service Application Pre and Post Test Procedures*.