Respironics Trilogy 100 Technical Information page 206

Table of Contents

Advertisement

8.5.6 T
100 F
RILOGY
Removal
1. Remove the all of the tubing, including the Transition Tube, from the Flow Sensor Assembly.
2. Remove the two screws securing the Flow Sensor Assembly to the Rear Enclosure.
3. Remove the Outlet Flow Path Thermistor and Temperature Sensor Caps from the Flow Sensor
Assembly.
4. Pull the Flow Sensor Assembly from the Rear Enclosure.
1002735, V
. 05
ER
S
A
LOW
ENSOR
SSEMBLY
F
8-44: S
IGURE
T
S
RILOGY
ERVICE
R
EPLACEMENT
L
CREW
OCATION
& T
I
ECHNICAL
NFORMATION
P
8-45
AGE