set to Flow Trigger, Flow Trigger Sensitivity and Flow Cycle Sensitivity become active, and the
trigger initiates based on the Flow Trigger Sensitivity setting.
Auto-Trak is only available if the Passive Circuit is selected.
Trigger Type is not available when Active PAP or Active Flow circuit is selected.
Flow trigger is the triggering method used for Active PAP and Active Flow circuits.
3. Flow Trigger Sensitivity
If you set the Trigger Type to Flow Trigger, the Flow Trigger Sensitivity setting displays. You can
increase or decrease the setting from 1 to 9 l/min in 1 l/min increments.
The flow trigger initiates when the patient's inspiratory effort creates a flow equal to or greater than
the flow sensitivity setting
1002735, V
. 05
ER
Flow Trigger with leak compensation is only available
if the Active Flow circuit is selected.