6.7 T
E
RILOGY
RROR
Use the corrective action(s) listed in section 6.5 (Troubleshooting Table) before using corrective
action(s) listed in the error code charts below.
6.7.1 V
ENTILATOR
ERROR
CODE
•
E-9
Motor Capacitor not installed
•
Motor Connector not soldered properly
•
Motor Rotor Locked or other mechanical
problem
•
E-14
Board revision resistors are not installed
•
Software is running on wrong revision of
the board
•
E-21
Problem
between DSP and Host
•
E-37
Host CPU is constantly rebooting
•
E-71
Parameter Settings corrupted
•
E-73
Parameter Settings corrupted
•
E-76
Defective EEPROM
E-77
•
Defective EEPROM
•
Program Execution Error
•
E-90
Software does not support newer/older
revision of Hardware
•
Board revision resistors on System/CPU
PCA are setup incorrect
•
E-101
Program Execution Error
•
E-103
Control Flow Sensor out of electrical spec
1002735, V
. 05
ER
C
ODES AND
I
A
NOPERATIVE
LARM
PROBABLE CAUSE
with
serial
communication
T
S
RILOGY
ERVICE
C
A
ORRECTIVE
CTIONS
NOTE
E
C
RROR
ODES
•
Check the Capacitor
•
Check Motor connector
•
Replace Motor
•
Replace System/CPU PCA
•
Load Correct Software
•
Replace System/CPU PCA
•
Replace System/CPU PCA
•
Reinstall Software
•
Replace System/CPU PCA
•
Replace System/CPU PCA
•
Replace System/CPU PCA
•
Replace System/CPU PCA
•
Replace System/CPU PCA
•
Reinstall software
•
Replace the System/CPU PCA
•
Contact Respironics Product Support
•
Check Tubing
•
Check Manifold
•
Replace Sensor PCA
& T
I
ECHNICAL
NFORMATION
CORRECTIVE ACTION
P
6-29
AGE