Respironics Trilogy 100 Technical Information page 110

Table of Contents

Advertisement

6.7 T
E
RILOGY
RROR
Use the corrective action(s) listed in section 6.5 (Troubleshooting Table) before using corrective
action(s) listed in the error code charts below.
6.7.1 V
ENTILATOR
ERROR
CODE
E-9
Motor Capacitor not installed
Motor Connector not soldered properly
Motor Rotor Locked or other mechanical
problem
E-14
Board revision resistors are not installed
Software is running on wrong revision of
the board
E-21
Problem
between DSP and Host
E-37
Host CPU is constantly rebooting
E-71
Parameter Settings corrupted
E-73
Parameter Settings corrupted
E-76
Defective EEPROM
E-77
Defective EEPROM
Program Execution Error
E-90
Software does not support newer/older
revision of Hardware
Board revision resistors on System/CPU
PCA are setup incorrect
E-101
Program Execution Error
E-103
Control Flow Sensor out of electrical spec
1002735, V
. 05
ER
C
ODES AND
I
A
NOPERATIVE
LARM
PROBABLE CAUSE
with
serial
communication
T
S
RILOGY
ERVICE
C
A
ORRECTIVE
CTIONS
NOTE
E
C
RROR
ODES
Check the Capacitor
Check Motor connector
Replace Motor
Replace System/CPU PCA
Load Correct Software
Replace System/CPU PCA
Replace System/CPU PCA
Reinstall Software
Replace System/CPU PCA
Replace System/CPU PCA
Replace System/CPU PCA
Replace System/CPU PCA
Replace System/CPU PCA
Reinstall software
Replace the System/CPU PCA
Contact Respironics Product Support
Check Tubing
Check Manifold
Replace Sensor PCA
& T
I
ECHNICAL
NFORMATION
CORRECTIVE ACTION
P
6-29
AGE