8.7.3 M
E
IXING
LEMENT
Removal
1. Remove the Filter/Whisper Cap. Refer to Oxygen Blender Filter/Whisper Cap Replacement Sec-
tion for more details.
2. Remove the Oxygen Blender Housing. Refer to the Oxygen Blender Housing Removal Section for
more details.
3. Release the Pressure Sensor Clip from the PCA.
4. Remove the rest of the tubing from the Flow Element Tubes.
5. Remove the screw securing the Mixing Element to the PCA.
6. Pull the Mixing element still connected to the Flow Element Tubes away from the Manifold.
7. Unscrew the screw securing the Flow Element Tubes to the Mixing Element.
Installation
1. Ensure that the two o-rings are in place at the end of the Flow Element Tubes.
2. Dip the end of the Flow Element Tubes in an alcohol bath and inert them into the Mixing Element.
3. Secure the Flow Element Tubes to the Mixing Element by securing the screw to 6 in-lbs.
4. Ensure that the two o-rings are in place at the other end of the Flow Element Tubes.
1002735, V
. 05
ER
/F
E
T
LOW
LEMENT
F
8-72: M
E
IGURE
IXING
LEMENT
T
S
RILOGY
ERVICE
R
UBES
EPLACEMENT
/F
T
R
LOW
UBES
EPLACEMENT
& T
I
ECHNICAL
NFORMATION
P
8-73
AGE